Welcome to the official website of Nanbei Group
NB-2800 Ion Chromatograph

NB-2800 Ion Chromatograph

Brand:NANBEI
Model:NB-2800 Ion Chromatograph
Application:Teaching research,environmental detection,optical material,metallic material,food safety,daily chemical products,agriculture,drug
Product Details
NB-2800 Ion Chromatograph


Introduction about 
Ion Chromatograph:
NB-2800 Ion Chromatography was created using the latest ion chromatography technology from both at home and abroad. With significant innovative breakthroughs, we were able to produce key components of the ion chromatography comparable to their international counterparts.From PEEKTM dual piston pumps and flow system to chemically-regenerated suppressors and eluent generators, NB-2800 with its technologically advanced components is a powerful analysis tool.Additionally with the Ace software, it is reliable, fast, and simple to use. With just some simple training, the NB-2800 can be used by even non-expert operators with all of its advanced capabilities.   

Main Features of 
Ion Chromatograph:
1. A chemically self-regenerating suppressor for a world class ion chromatograph.
When analyzing the eluent with a high background conductivity, the ion chromatograph is unable to receive a signal from the analyte through the chemical suppressor. To address this problem, when analyzing anions, CO32- and HCO3- is added to the eluent. By generating H+ ions through electrolysis, a low background conductivity can be fostered by forming H2CO3.When analyzing cations, H+ ions generated by electrolyzing water reacts with OH- ions to form water, and, therefore, decrease the background conductivity. Using electrolysis to produce H+ or OH- will also chemically regenerate the ion-exchange membranes.   

Suppressor (Anionic/Cationic):
We provide two kinds of chemically-regenerated supressors: anionic and cationic. We are the first in China to commericially produce anionic chemical suppressors. Our suppressor can handle large amounts and has a low background conductivity (able to detect up to ppb concentrations). Additionally, it has a small dead volume, good balance, and high repeatability. It is easy-to-use with almost no maintenance needed.

2. PEEKTM dual pistons and a low pulse infusion pump for smooth operation and low maintenance costs.

3. Advanced digital conductivity detector with high stability and high sensitivity for more accurate and reliable results.

4. PEEKTM flow system able to withstand high pressures, resist corrosion by acids and akali, prevent metal contamination, and is compatible with organic solvents.

5. Fast data transmissions and processing. Automatically identifies sample while controlling and monitoring the instrument in real-time to guarantee a continuous and stable analysis.

6. Optional eluent generator to automatically prepare eluents for ion chromatography and save a significant amount of time and effort.

7. Optional Auto-sampler

8. Software
A powerful, easy-to-use Ace chromatography software. With only a few clicks of the mouse or by using our intuitive front panel interface, you can maintain complete control of the instrument. Using the real-time monitoring system, you can save time and work through the entire analysis process.


Advantages/Benefits of 
Ion Chromatograph:

1. Structure
Easy to maintain and to use instrument, especially the infusion pumps and basic components of the detector

2. PEEKTM Infusion Pump
a) Ion chromatography can analyze both acidic and basic eluents. Consequently, the flow path must be corrosion-resistant, able to withstand high pressures, and remain unaffected by chemical reagents. Our PEEKTM Infusion Pump seeks to address these needs for maximum efficiency and with a long working lifespan. It is capable of withstanding pressures of up to 35 MPa with a maximum flow rate of 5 mL/min.

b) A conductivity detector is used as an universal detector for an ion chromatography. Because the solution flow rate must be highly stable, our infusion pump currently uses a new buffer tube. The buffer greatly reduces the number of pressure fluctuations in the pump, which increases the flow accuracy to 0.2 %. This eliminates baseline shifts caused by an unstable flow rate.

c) Automatic cleaning: Ensures maximum working efficiency by preventing the pump from drying up or from forming salt crystals. This limits any damage to the piston ring and the piston rods, which extends the pump's lifespan and reduces operating costs.

3. Chemically Self-Regenerating Suppressor
a) An internal structure specifically designed to reduce dead volume (< 50ml), as a large dead volume can cause a broad peak.

b) A platinum plated titanium electrode acts as a supportive membrane and as a flow path. In contrast with an insulated plate's, the titanium electrode's membrane has a greatly reduced voltage, and, thus, it prevents high voltages caused by thermal interference.

c) No need for strong acids or akalis to regenerate the membrane. Instead, through the electrolysis of water H+ and HO- ions are then used to regenerate the membrane.As such, the instrument is safe and easy to operate.

4. Conductivity Conductor
a) Because the conductivity of a solution is affected by temperature, we use a conductivity conductor with a temperature control system. With a temperature control accuracy of 0.01 °C, it is able precisely determine conductivity without interference caused by temperature.

b) A dynamic fluid testing structure significantly improves detection repeatability. Results from road tests indicate that the detector is unaffected to motion changes.

5. Operating Software
a) This Windows based operating software can maintain complete control of the instrument and monitor instrument status in real-time. With just a few clicks of the mouse, you can complete a full analysis.

b) RS232 and an optional USB interface for greater flexibility and convenience.

6. EG100 Automatic Eluent Generator
a) EG100 is able to generate all kinds of eluents and single pump concentration gradients using a well-designed internal structure without a degassing device. With a longer working lifespan, it is responsible for more reliable and more secure analysis.

b) EG100 is compatible with a wide range of ion chromatographs. By attaching it to any ion chromatograph, the ion chromatograph will only require water to be used.

Technical Specifications

Detectable Ions

Anions: F-, Cl-, NO2-, Br-, NO3-, HPO42-, SO32-, S2O32-, SO42-, formic acid, acetic acid, oxalic acid, water disinfectant byproducts
Cations: Li+, Na+, NH+, K+, Mg2+, Ca2+
Detection range: ppb to ppm level
Detection Limit: 0.1 μg/L (Cl-) 5.0 μg/L (Na+)
Linear Dynamic Range: 103
Linear Correlation Coefficient: 0.9998 (Cl- and Na+)
Baseline Noise: ≤ 0.05mV/30min
Baseline Drift for Anions: ≤ ± 1.5mV/30min
Baseline Drift for Cations: ≤ ± 0.5mV/30min


Infusion Pump:
Type: Parallel Dual-Piston Reciprocating Pump, microprocessor controlled fixed strokes, adjustable speed
Structure: Chemically inert, non-metallic PEEKTM pump head and flow path
PH: 0-14
Control Modes: Full control through Ace software or use of front panel
Pressure: maximum 35 MPa (5000 psi)
Flow Rate: 0.01 to 5.0 ml/min, in increments of 0.01
Flow Accuracy: 0.1% RSD
Flow Accuracy: ± 2%
Piston Cleaning: Automatic cleaning for dual-piston
Overvoltage Protection: up to 0-35 MPa in 1 V increments, the lower limit can be set as 1 V lower than the upper limit, automatic pump shutdown when limits are exceeded
Degasser (optional): Dual-channel, automated online.

Conductivity Detector:
Type: Microprocessor controlled digital signal processing
Cell Drive: 10 kHz
Output Range: 0 - 15000 μS/cm
Cell Temperature: Room temperature to -60 ℃, user settable
Cell Temperature Stability: ≤ 0.005 °C
Cell Body: PEEKTM
Flow Cell Volume: < 1μL
Flow Cell Maximum Pressure: 3 MPa (450 psi)

Suppressor:
Modes of Operation Supported: Chemical regeneration
Suppression Amount for Anion: 100 mmol NaOH
Suppression Amount for Cation: 100 mmol MSA
Time to Reach Equilibrium State: < 15 min
Current (Anions): 0 - 200 mA, 1 mA increments
Current (Cations): 0 - 300 mA in 1 mA increments

Eluent Generator:
Minimum and Maximum Eluent Concentrations: 0.1 - 50 mmol/L
Eluent Types: OH-, CO32-/HCO3-,MS
Incremental concentration: 0.1 mmol/L
Flow Rate: 0.5 - 3.0 mL/min
Temperature: Room temperature - 40 ℃
Humidity: 5% - 85% relative humidity, no condensation
Dimensions (L × W × H): 586mm × 300mm × 171mm
Weight: 5 kg

General Specifications:
Power Supply: 220 ±10 V, 50/60Hz
Temperature: 5 ℃ - 40 ℃
Humidity: 5% - 85% relative humidity, no condensation
Ports: RS485 (USB optional)
Dimensions (L × W × H): 586mm × 300mm × 350mm
Weight: 34kg
Power: 150W
Product title:
*E-mail:
The other contacted ways:
Product model:
*Detail requirements: