 AA-7050 Series Atomic Absorption Spectrometer

AA-7050 Series Atomic Absorption Spectrometer

Brand:NANBEI
Model: AA-7050 Series Atomic Absorption Spectrometer
Application:Teaching research,environmental detection,optical material,metallic material,food safety,daily chemical products,agriculture,drug
Product Details
AA-7050 Series Atomic Absorption Spectrometer


Introduction about  AA-7050 Series Atomic Absorption Spectrometer:
* The graphite furnace camera helps users locate injection probe, view sample status in graphite tubes and optimize furnace temperature programming. 
* Graphite furnace gas saving mode prolongs the usage time of protection gas and therefore reduces the operating cost of the instrument.
* Graphite furnace double inner gas path permits an auxiliary gas in addition to the protection gas so as to improve the ashing process and obtain better results
* 6 coded lamp turret can automatically identify lamp element and record element lamp usage times.
* Deuterium lamp fine adjustment mechanism ensures perfect overlap of deuterium lamp and element lamp positions.
* Multi-element analysis function and AS-600 flame/graphite furnace auto sampler are equipped to realize fully automated analysis of multiple elements

 AA-7050 Series Atomic Absorption Spectrometer Specifications:
 
1. Optics System:
 a)  Wavelength Range: 190nm~900nm
 b)  Band Width: 0.1, 0.2, 0.4, 1.0 and 2.0nm, automatic slit switch
 c)  Mono-Chromator: C-T grating Mono-Chromator
 d)  Wavelength accuracy/ error: ±0.1nm
 e)  Wavelength repeatability: ≤0.05nm
 f)  Optical grating: 1800 lines/mm
 g)  Blaze wavelength: 250nm
 h)  Baseline stability: static state baseline≤0.002A/30min, dynamic baseline≤0.003A/30min
 i)  Resolution: better than 0.1nm
 
2. Flame Method:
 a)  Characteristic concentration: ≤0.02 μg/mL/1% (Cu)
 b)  Detection limit: ≤0.003 μg/ml
 c)  RSD≤0.45%
 d)  Burner head: 50mm and 100mm interchangeable full titanium burner head and full titanium nebulizer.
 e) Position adjustment: adjustable height and angles. Flame/hydride method can be switched in 1 minute.
 
3. Graphite Furnace Method:
 a)  Characteristic concentration: ≤0.3×10-12g (Cd)
 b)  Detection limit: ≤0.2×10-12g
 c)  RSD≤1.8%
 d)  Temperature control program: Max 20 ramp temperature increasing programming in step, slope or flat mode.
 e) Temperature increasing by power or optical control
 f)  Temperature increasing rate: ≥2000℃/s by power, ≥3000℃/s by optical control.
 
4. Background Calibration:
 a) Both flame and furnace method can run background calibration. b) Background: deuterium lamp, self absorption (optional)
 c) Calibration capability: AAS has the calibration capability of more than 60 times when background absorption value is close to 1.0Abs
 
5. Data Processing:
 a)  Measurement methods: flame absorption, flame emission, graphite furnace and hydride.
 b)  Analysis methods: linear equation, non-linear equation, added standard.
 c)Print output: calibration curve, signal spectrogram, instrument conditions, analysis parameters and analysis results can all be automatically saved and printed.
 
6. Specifications:
 a)  Dimensions: 890mm (L)×567mm (W)×490mm (H), 125kg
 b)  Power: 220V 50Hz one-phase power, main unit 200W, graphite furnace 4KW.

 